Please observe the following points:


-
In order to avoid incorrect (too high) CO values from being displayed, check the Purafil filter cartridge prior to each measurement. As soon as the first third of the violet-red granulate turns grayish brown or when moisture becomes visible (dark granulates), the whole cartridge or the granulate must be replaced.
-
When filling the cartridge, replace both synthetic fiber wads. Only use artificial fiber wads.
-
When filling with granulate, place cartridge on hard surface, tap and fill to the top.
-
When opening the cartridge, unscrewing the caps avoids damaging the O-rings.
-
Correct handling prevents faults and the operating life of the sensors is extended.
-
After each measuring operation, the water separator must be emptied and kept open, allowing them to dry out. Leaking condensate may damage the analyzer. Caution: do not loose seral.
-
The hoses may not be kinked.
-
Regularly clean the gas extraction probe and hoses.

Important


The analyzer should be revised by an anapol approved service location at least once a year or every 1000 operating hours.
If the analyzer is only used during a certain period, the analyzer must be revised prior to this period.

Please read the detailed information in the users manual regarding the maintenance of the device.